Repeatability of swept-source optical coherence tomography angiography automated macular vessel density

Fecha de publicación: Fecha Ahead of Print:

Autores de FISABIO

Autores ajenos a FISABIO

  • Monferrer-Adsuara, C
  • Navarro-Palop, C
  • Cervera-Taulet, E
  • Montero-Hernandez, J
  • Medina-Besso, P
  • Castro-Navarro, V

Grupos de Investigación

Abstract

Introduction: Optical Coherence Tomography Angiography Ratio Analysis (OCTARA) is capable of visualizing inner and outer retinal vascular plexuses, choriocapillaris, and larger choroidal vasculature in vivo without contrast injection. The aim of this study was to assess the intrasession repeatability of automated vessel density measurements using Triton Swept-Source Optical Coherence Tomography Angiography (OCTA) innovative algorithm OCTARA in retinal and choroidal vasculature. Methods: To study population between 65-90 years old with no eye diseases. For each subject measurements were performed four times. The intraclass correlation coefficient and the coefficient of variation were calculated to analyze repeatability of the OCTARA automatically generated vessel density measurements. Results: A total of 35 eyes were included in the study. The intraclass correlation coefficient of the global vessel density in the superficial capillary plexus and the deep capillary plexus were 0.963 and 0.975, respectively, and the coefficient of variation were 5.4% and 4.4%, respectively. The intraclass correlation coefficient of the rest of the global measurements was indicative of good reliability with the exception of the deep choroid layer with an intraclass correlation coefficient of 0.6 indicative of moderate reliability. Conclusions: Our results proved excellent repeatability of automated vessel density measurements in the superficial and deep capillary plexus layers in our cohort using a OCTARA algorithm indicating that it may be a reliable diagnostic tool. It also showed good reliability in choriocapillary and mid choroid layer. These findings may be of value in assessing the significance of differences in capillary density measurements over time and across different settings.

Datos de la publicación

ISSN/ISSNe:
1120-6721, 1724-6016

EUROPEAN JOURNAL OF OPHTHALMOLOGY  SAGE PUBLICATIONS LTD

Tipo:
Article
Páginas:
1964-1974
PubMed:
38374527

Citas Recibidas en Web of Science: 3

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Anatomy/biochemistry/physiology < RETINA; telemedicine < SOCIOECONOMICS AND EDUCATION IN MEDICINE/OPHTHALMOLOGY; practice management < SOCIOECONOMICS AND EDUCATION IN MEDICINE/OPHTHALMOLOGY; preventive medicine/screening < SOCIOECONOMICS AND EDUCATION IN MEDICINE/OPHTHALMOLOGY; techniques of retinal examination < RETINA

Cita

Compartir